QC Procedure
Using Igloo Reader Pro with QC Calibration Test
π¦ Step 1:
Prepare the QC Cartridge
- Make sure you have an official QC calibration cassette (purchased separately)
- Insert the cassette into the designated QC adapter
- Ensure the cassette is undamaged and not expired
π Step 2:
Power on the Reader
- Press and hold the round power button located on the bottom side of the Igloo Reader Pro
- Wait for the screen to fully load to the main menu
π Step 3:
Insert the Adapter into the Reader
- Place the adapter with the QC cassette into the reader slot
- Make sure it is fully inserted and properly aligned
βοΈ Step 4:
Automatic QC Detection
- The reader will automatically recognize the QC process
- No manual action is required β the system will start the validation check
β³ Step 5:
Wait for Result
- The display will show:
- β QC successful β validity extended for 12 months
- β QC failed β repeat test or contact technical support
β οΈ Do not remove the adapter during this step. Removing it will cancel the test.
π Step 6:
Confirm QC Status
- To verify the QC expiration:
- Go to Settings β About
- Check the field: QC valid until [date]